1887
banner image
No data available.
Please log in to see this content.
You have no subscription access to this content.
No metrics data to plot.
The attempt to load metrics for this article has failed.
The attempt to plot a graph for these metrics has failed.
Temporal and spectral imaging with micro-CT
Rent:
Rent this article for
USD
10.1118/1.4736809
/content/aapm/journal/medphys/39/8/10.1118/1.4736809
http://aip.metastore.ingenta.com/content/aapm/journal/medphys/39/8/10.1118/1.4736809

Figures

Image of FIG. 1.
FIG. 1.

A diagram of the proposed technique.

Image of FIG. 2.
FIG. 2.

The material decomposition can be represented as a linear transformation from attenuation space (b) and (e) to material space (c) and (f). The attenuation coefficients for water and iodine are represented by vectors, and the axes of the ellipses around the vectors are the standard deviations of these measurements in a typical scan with our micro-CT system (b), (e). These correspond to the vectors and ellipses representing the relative concentrations (0–1) of water and iodine (c) and (f). When the energy settings are suboptimal (a), the vectors corresponding to water and iodine are close together (b), and the transformation amplifies the noise (c). When a better set of energy settings are chosen (d), the water and iodine vectors are farther apart (e), and the noise amplification is less pronounced (f). The condition number of the decomposition matrix corresponding to the improved settings is less than half of the condition number for the original settings.

Image of FIG. 3.
FIG. 3.

The condition number of the material decomposition matrix as a function of voltage, without (a) and with (b) filters. The darkest region in B, 40 kVp with the tin filter and 55 kVp with the tungsten filter, indicated with the crosshairs, corresponds to the lowest condition number.

Image of FIG. 4.
FIG. 4.

The spectral sensitivity of our system with the two x-ray energy settings. These functions include the contributions from the tube, filters, and detector.

Image of FIG. 5.
FIG. 5.

Spectral basis functions.

Image of FIG. 6.
FIG. 6.

Typical temporal basis functions and sensitivity function.

Image of FIG. 7.
FIG. 7.

Examples of spectral-temporal sensitivity functions (a) and basis functions (b). These functions are normalized and dimensionless.

Image of FIG. 8.
FIG. 8.

Reconstructed attenuation images from the simulated scan using both original projections and interpolated projections, without and with bilateral filtration. Images with the low energy setting at two cardiac phases are shown.

Image of FIG. 9.
FIG. 9.

Measures of reconstruction quality at each iteration of the refinement process, starting with the result from the material decomposition (iteration 0): The RMSE of the estimated iodine concentration at all points in the iodine volumes at all points in time (a), the STD (b) and MTF (c) measured around the vial containing water, and the SE of the left ventricle (d).

Image of FIG. 10.
FIG. 10.

Images of the water and iodine components at one cardiac phase, before and after iterative refinement.

Image of FIG. 11.
FIG. 11.

Reconstructed images from the in vivo scan. An attenuation image reconstructed with all the projections is shown at the top. The material components at two cardiac phases are shown below, with water in grayscale and iodine in red.

Image of FIG. 12.
FIG. 12.

A closer view of a coronal section of the iodine in the heart over eight phases of the cardiac cycle, with the same color map as in Fig. 11.

Image of FIG. 13.
FIG. 13.

The volume of the left ventricle of the heart over eight phases of the cardiac cycle, calculated from segmented images.

Tables

Generic image for table
TABLE I.

Measurements of the reconstructed attenuation images from the simulation, with sorted projections and interpolated projections, without and with bilinear filtration.

Generic image for table
TABLE II.

Computation time of the steps of the reconstruction process for the in vivo study.

Loading

Article metrics loading...

/content/aapm/journal/medphys/39/8/10.1118/1.4736809
2012-07-25
2014-04-19
Loading

Full text loading...

This is a required field
Please enter a valid email address
752b84549af89a08dbdd7fdb8b9568b5 journal.articlezxybnytfddd
Scitation: Temporal and spectral imaging with micro-CT
http://aip.metastore.ingenta.com/content/aapm/journal/medphys/39/8/10.1118/1.4736809
10.1118/1.4736809
SEARCH_EXPAND_ITEM