Imaging dose in variable pitch body perfusion CT scans: An analysis using TG111 formalism
Rent:
Rent this article for
Access full text Article
/content/aapm/journal/medphys/41/6/10.1118/1.4876377
1.
1. A. R. Kambadakone and D. V. Sahani, “Body perfusion CT: Technique, clinical applications, and advances,” Radiol. Clin. North Am. 47(1), 161178 (2009).
http://dx.doi.org/10.1016/j.rcl.2008.11.003
2.
2.State of California, Radiation control: Health facilities and clinics: Records, Bill number AB-510, 2012, see http://leginfo.legislature.ca.gov/faces/billVotesClient.xhtml.
3.
3. J. M. Boone, “The trouble with CTD100,” Med. Phys. 34(4), 13641371 (2007).
http://dx.doi.org/10.1118/1.2713240
4.
4. R. L. Dixon and A. C. Ballard, “Experimental validation of a versatile system of CT dosimetry using a conventional ion chamber: Beyond CTDI100,” Med. Phys. 34(8), 33993413 (2007).
http://dx.doi.org/10.1118/1.2757084
5.
5.Safety Investigation of CT Brain Perfusion Scans, 2010, see http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm185898.htm.
6.
6.Adult Brain Perfusion CT Protocols, Version 1.1, 2012, see http://www.aapm.org/pubs/CTProtocols/documents/AdultBrainPerfusionCT.pdf.
7.
7. AAPM, Task Group Report No. 111, Comprehensive Methodology for the Evaluation of Radiation Dose in X-Ray Computed Tomography, 2010.
8.
8. AAPM, Task Group Report No. 96, The Measurement, Reporting, and Management of Radiation Dose in CT, 2008.
9.
9. W. Feng, C. Schultz, H. Chen, A. Chu, A. Hu, and D. Bakalyar, “TU-FF-A4-04: Experimental confirmation of near parabolic shape of dose profile in cylindrical phantom for dual source CT,” Med. Phys. 34(6), 25712572 (2007).
http://dx.doi.org/10.1118/1.2761440
10.
10. Y. Liang, A. Dutta, S. Jambawalikar, J. So and D. Lynch, “TU-G-103-09: Measurement of Planar Average Equilibrium Dose of CT,” Med. Phys. 40(6), 460 (2013).
http://dx.doi.org/10.1118/1.4815478
11.
11. S. Mutic, J. R. Palta, E. K. Butker, I. J. Das, M. S. Huq, L.-N. D. Loo, B. J. Salter, C. H. McCollough, and J. Van Dyk, “Quality assurance for computed-tomography simulators and the computed-tomography-simulation process: Report of the AAPM Radiation Therapy Committee Task Group No. 66,” Med. Phys. 30(10), 27622792 (2003).
http://dx.doi.org/10.1118/1.1609271
12.
12. International Electrotechnical Commission, “Medical Electrical Equipment—Part 2-44: Particular requirements for the basic safety and essential performance of X-ray equipment for computed tomography,” IEC60601-2-44 (International Electrotechnical Commission, Geneva, Switzerland, 2009).
13.
13. R. L. Dixon and J. M. Boone, “Cone beam CT dosimetry: A unified and self-consistent approach including all scan modalities—With or without phantom motion,” Med. Phys. 37(6), 27032718 (2010).
http://dx.doi.org/10.1118/1.3395578
http://aip.metastore.ingenta.com/content/aapm/journal/medphys/41/6/10.1118/1.4876377
Loading
/content/aapm/journal/medphys/41/6/10.1118/1.4876377
Loading

Data & Media loading...

Loading

Article metrics loading...

/content/aapm/journal/medphys/41/6/10.1118/1.4876377
2014-05-21
2014-08-23

Abstract

To investigate the variation of imaging dose with tube potential in variable pitch body CT perfusion (CTp) protocols using the TG111 dosimetric formalism.

TG111 recommendations were followed in choosing the phantom, dosimetric equipment, and methodology. Specifically, equilibrium doses ( ) were measured centrally and peripherally in a long PMMA phantom. Reference planar average equilibrium doses were determined for each tube potential, for a reference set of exposure parameters (collimation, pitch, filtration) on a Siemens Definition CT scanner. These reference values were utilized to predict the imaging dose during perfusion scans using interpretations of the TG111 formalism. As a gold reference, the midscan average planar perfusion doses ( ) were obtained directly from central and peripheral measurements for body CTp scans (144 and 271 mm) using variable pitch acquisition. Measurement-based values obtained using a thimble chamber were compared to the TG111-predicted values, and to CTDI reported at the console.

Reference planar average equilibrium dose values measured for reference uniform pitch helical scans were consistently higher than console-reported or measured values for CTDI . The measurement-based perfusion dose was predicted accurately by the reported CTDI for the 144 mm scan. The 271 mm scans delivered systematically larger dose than reported. The TG111-based dose estimates were proven to be conservative, as they were systematically higher than both the measured and the reported imaging doses.

Upon successful implementation of TG111 formalism, standard imaging dose was measured for a body CTp protocol using the variable pitch helical acquisition. The TG111 formalism is not directly applicable to this type of acquisition. Measurement of dose for all variable pitch protocols is strongly suggested.

Loading

Full text loading...

/deliver/fulltext/aapm/journal/medphys/41/6/1.4876377.html;jsessionid=cls6ge5t5ftlq.x-aip-live-02?itemId=/content/aapm/journal/medphys/41/6/10.1118/1.4876377&mimeType=html&fmt=ahah&containerItemId=content/aapm/journal/medphys

Most read this month

Article
content/aapm/journal/medphys
Journal
5
3
Loading

Most cited this month

true
true
This is a required field
Please enter a valid email address
This feature is disabled while Scitation upgrades its access control system.
This feature is disabled while Scitation upgrades its access control system.
752b84549af89a08dbdd7fdb8b9568b5 journal.articlezxybnytfddd
Scitation: Imaging dose in variable pitch body perfusion CT scans: An analysis using TG111 formalism
http://aip.metastore.ingenta.com/content/aapm/journal/medphys/41/6/10.1118/1.4876377
10.1118/1.4876377
SEARCH_EXPAND_ITEM