Cone-beam computed tomography with a flat-panel imager: Initial performance characterization

Med. Phys. Volume 27, Issue 6, pp. 1311-1323 (June 2000)

Issue Date: June 2000
Buy This PDF   (US$28)
Download PDF (1572 kB) View Cart

KEYWORDS and PACS

Keywords
PACS
  • 87.59.Fm
    Biological and medical physics X-ray imaging Computed tomography (CT)
  • 87.56.Da
    Biological and medical physics Radiation therapy equipment Ancillary equipment
  • 87.62.+n
    Biological and medical physics Medical imaging equipment
  • 42.79.Kr
    Optics Optical elements, devices, and systems Display devices, liquid-crystal devices
  • YEAR: 2000

PUBLICATION DATA

ISSN:
0094-2405 (print)  
Publisher:
AIP is a member of CrossRef AAPM
D. A. Jaffray and J. H. Siewerdsen
Department of Radiation Oncology, William Beaumont Hospital, Royal Oak, Michigan 48073
The development and performance of a system for x-ray cone-beam computed tomography (CBCT) using an indirect-detection flat-panel imager (FPI) is presented. Developed as a bench-top prototype for initial investigation of FPI-based CBCT for bone and soft-tissue localization in radiotherapy, the system provides fully three-dimensional volumetric image data from projections acquired during a single rotation. The system employs a 512×512 active matrix of a-Si:H thin-film transistors and photodiodes in combination with a luminescent phosphor. Tomographic imaging performance is quantified in terms of response uniformity, response linearity, voxel noise, noise-power spectrum (NPS), and modulation transfer function (MTF), each in comparison to the performance measured on a conventional CT scanner. For the geometry employed and the objects considered, response is uniform to within 2% and linear within 1%. Voxel noise, at a level of ~20 HU, is comparable to the conventional CT scanner. NPS and MTF results highlight the frequency-dependent transfer characteristics, confirming that the CBCT system can provide high spatial resolution and does not suffer greatly from additive noise levels. For larger objects and/or low exposures, additive noise levels must be reduced to maintain high performance. Imaging studies of a low-contrast phantom and a small animal (a euthanized rat) qualitatively demonstrate excellent soft-tissue visibility and high spatial resolution. Image quality appears comparable or superior to that of the conventional scanner. These quantitative and qualitative results clearly demonstrate the potential of CBCT systems based upon flat-panel imagers. Advances in FPI technology (e.g., improved x-ray converters and enhanced electronics) are anticipated to allow high-performance FPI-based CBCT for medical imaging. General and specific requirements of kilovoltage CBCT systems are discussed, and the applicability of FPI-based CBCT systems to tomographic localization and image-guidance for radiotherapy is considered. ©2000 American Association of Physicists in Medicine.
History: Received 15 September 1999; accepted 29 March 2000
Permalink: http://dx.doi.org/10.1118/1.599009