Analyzer system for field emission energy distribution (FEED) measurements
Rev. Sci. Instrum. 49, 1285 (1978); doi:10.1063/1.1135569
Issue Date: September 1978
You are logged in to this journal.
A newly designed analyzer system for measuring the energy distribution of field emitted electrons (FEED) is presented. A Kuyatt–Simpson type lens system is combined with a two-stage 127° analyzer. A resolution of 15 meV is obtained with a transmission energy of 0.5 eV.
Review of Scientific Instruments is copyrighted by The American Institute of Physics.
| Permalink: |
http://link.aip.org/link/?RSINAK/49/1285/1 |
REFERENCES (11)
-
J. W. Gadzuk and E. W. Plummer, Rev. Mod. Phys. 45, 487 (1973).
-
R. D. Young and E. W. Müller, Phys. Rev. 113, 115 (1959). [ISI]
-
A. van Oostrom, Philip Res. Rep. Suppl. 11, 102 (1966).
-
C. E. Kuyatt and J. A. Simpson, J. Appl. Phys. 37, 3805 (1966); [ISI]
-
C. E. Kuyatt and E. W. Plummer, Rev. Sci. Instrum. 43, 108 (1972). [ISI]
-
K. Klapper, Ph.D. thesis (Universität München, 1978).
-
Ch. Todd, Ph.D. thesis (Cornell, Ithaca, NY, 1971).
-
R. D. Young and C. E. Kuyatt, Rev. Sci. Instrum. 39, 1477 (1968). [ISI]
-
C. Lea and R. Gomer, Phys. Rev. Lett. 25, 804 (1970). [ISI]
-
N. J. Dionne and T. N. Rhodin, Phys. Rev. Lett. 32, 1311 (1974);
-
O. K. Andersen, Phys. Rev. B 2, 883 (1970). [Inspec]