X-ray monochromator with an energy resolution of 8×109 at 14.41 keV
Rev. Sci. Instrum. 72, 4080 (2001); doi:10.1063/1.1406925
Issue Date: November 2001
You are not logged in to this journal. Log in
An ultrahigh-resolution x-ray crystal monochromator providing a 120 µeV bandwidth at 14.41 keV is presented. The design, which uses four independent silicon crystals and has an output beam parallel to the incident beam, may be generalized to arbitrary energies. Fluxes of 1.3×106 photons/s (1.0×107 photons/s) in bandwidths of 120±15 µeV (140±15 µeV) were measured. The performance of the monochromator, including the preservation of coherence through it, was verified by measurement of a 9.6±2.0% enhancement in the coincidence rate (i.e.,
(2)1 = 0.096±0.020) in an intensity correlation experiment. ©2001 American Institute of Physics.
(2)1 = 0.096±0.020) in an intensity correlation experiment. ©2001 American Institute of Physics.
| History: | Received 30 March 2001; accepted 6 August 2001 |
| Permalink: |
http://link.aip.org/link/?RSINAK/72/4080/1 |
REFERENCES (28)
For access to fully linked references, you need to log in.
For access to fully linked references, you need to Log in.
- Nuclear Resonant Scattering of Synchrotron Radiation, edited by E. Gerdau and H. de Waard (Baltzer Science, Bussum, The Netherlands, 1999/2000).
- See, for example, E. Burkel, Inelastic Scattering of X-rays with Very High Energy Resolution (Springer, Berlin, 1991);
- F. Sette, M. H. Kirsch, C. Masciovecchio, G. Ruocco, and G. Monaco,
Science 280, 1550 (1998) , and references therein. - W. Graeff and G. Materlik,
Nucl. Instrum. Methods Phys. Res. 195, 97 (1982) ; - B. Dorner, E. Burkel, and J. Peisl,
Nucl. Instrum. Methods Phys. Res. A 426, 450 (1986) ; - R. Verbeni et al.,
J. Synchrotron Radiat. 3, 62 (1996) . - G. Faigel, D. P. Siddons, J. B. Hastings, P. E. Haustein, J. R. Grover, J. P. Remeika, and A. S. Cooper, Phys. Rev. Lett. 58, 2699 (1987).
- T. Ishikawa, Y. Yoda, K. Izumi, C. K. Suzuki, X. W. Zhang, M. Ando, and S. Kikuta, Rev. Sci. Instrum. 63, 1015 (1992).
- T. S. Toellner, T. S. Mooney, S. Shastri, and E. E. Alp,
Proc. SPIE 1740, 218 (1992) . - D. P. Siddons, U. Bergmann, and J. B. Hastings, Phys. Rev. Lett. 70, 359 (1993).
- A. I. Chumakov, J. Metge, A. Q. R. Baron, H. Grünsteudel, H. F. Grünsteudel, R. Rüffer, and T. Ishikawa,
Nucl. Instrum. Methods Phys. Res. A 383, 642 (1996) . - T. S. Toellner, M. Y. Hu, W. Sturhahn, K. Quast, and E. E. Alp, Appl. Phys. Lett. 71, 2112 (1997).
- A. I. Chumakov, R. Rüffer, O. Leupold, A. Barla, H. Thiess, T. Asthalter, B. P. Doyle, A. Snigirev, and A. Q. R. Baron, Appl. Phys. Lett. 77, 31 (2000).
- K. Kohra and T. Matsushita,
Z. Naturforsch. Teil A 27A, 484 (1972) . - S. Kimura, J. Harada, and T. Ishikawa,
Acta Crystallogr., Sect. A: Found. Crystallogr. 50, 337 (1994) . - We note that A. Q. R. Baron et al. recently developed a compact in-line HRM with four-bounce reflections including Bragg angles ~
/2, where these problems are sophisticatedly solved. [A. Q. R. Baron, Y. Tanaka, D. Ishikawa, D. Miwa, M. Yabashi, A. I. Chumakov, and T. Ishikawa,
J. Synchrotron Radiat. 8, 1127 (2001)]. - K. Kohra and K. Kikuta,
Acta Crystallogr., Sect. A: Cryst. Phys., Diffr., Theor. Gen. Crystallogr. 24, 200 (1968) ;
T. Ishikawa, - J. W. M. DuMond, Phys. Rev. 52, 872 (1937);
- This energy was chosen for convenience in measuring the bandwidth via scattering from the nuclear resonance in 57Fe.
- S. Kikuta and K. Kohra,
J. Phys. Soc. Jpn. 29, 1322 (1970) . - M. Yabashi et al.,
Nucl. Instrum. Methods Phys. Res. A 467468, 678 (2001) . - J. B. Hastings, D. P. Siddons, U. van Bürck, R. Hollatz, and U. Bergmann, Phys. Rev. Lett. 66, 770 (1991).
- In general, the resolution measurement using NFS with crystal rotation is valid when the energy spectrum of the incident beam for the rotated crystal (fourth crystal in our case) is sufficiently larger than the spectrum of the exit beam from the crystal. For our case, this condition was verified not only by the DuMond diagram analysis theoretically (Fig. 2), but also by the rocking curve measurements experimentally, where the rocking curve width for nonresonant signal (~6 µrad) was sufficiently larger than that for NFS signal (~1 µrad).
- Incident flux without the slit is estimated to be 6×1013 photon/s with a beam size of 0.5(V)×1.3 (H) mm2 (FWHM) at the monochromator position. Apart from the decrease of the throughput discussed in the text, the throughput was reduced by a factor of ~15 because of the spatial acceptance of the HRM limited in the horizontal direction.
- For the calculation, a horizontal angular source size of 13.0 µrad (FWHM) seen from the UHRM position is used as an incident angular divergence, because the UHRM spatial acceptance is sufficiently smaller than the horizontal source size, 0.87 mm (FHWM).
- A. Q. R. Baron, Y. Kohmura, Y. Ohishi, and T. Ishikawa, Appl. Phys. Lett. 74, 1492 (1999);
- E. Ikonen, Phys. Rev. Lett. 68, 2759 (1992).
- Y. Kunimune, Y. Yoda, K. Izumi, M. Yabashi, X. W. Zhang, T. Harami, M. Ando, and S. Kikuta,
J. Synchrontron Radiat. 4, 199 (1997) ;
E Gluskin, E. E. Alp, I. McNulty, W. Sturhahn, and J. Sutter, - R. Hanbury-Brown and R. Q. Twiss,
Nature (London) 177, 27 (1956) . - During the experiment the SPring-8 storage ring was operated in 406 bunch mode with a bunch interval of 11.8 ns, which is sufficiently larger than the time resolution of the coincidence unit (Phillips Scientific Research, 754) used here. Actually, 32 buckets of 406 buckets were operated as empty bunches in order to avoid instability of orbital electrons, although this did not affect our experiment.
- L. Mandel and E. Wolf, Optical Coherence and Quantum Optics (Cambridge University Press, Cambridge, 1995).